Protective effect of intraoperative nerve monitoring against recurrent laryngeal nerve injury during re-exploration of the thyroid

Abstract

Background

Previous thyroid or parathyroid surgery induces scarring or distorts anatomy, and increases the risk of recurrent laryngeal nerve (RLN) injury for a reoperation. The benefit of intraoperative nerve monitoring (IONM) for re-exploration (a second nerve exploration) and reoperation has not been established.

Methods

Two hundred and ten patients were given a thyroid or parathyroid reoperation at our hospital between 2001 and 2010. Using IONM, we re-explored 56 patients who had been operated on before June 2007. The injury rate in these patients was compared with that of the 15 patients re-explored without IONM between 2001 and 2006.

Results

Of the 70 nerves that were re-explored using IONM, only one was incidentally injured, significantly fewer than the three injured in the 15 nerves re-explored without using IONM (1.43% vs. 20%, P = 0.0164).

Conclusions

IONM helped prevent RLN damage when re-exploring nerves during thyroid and parathyroid surgery. We recommend the routine use of IONM in thyroid and parathyroid reoperations.

Background

Recurrent laryngeal nerve (RLN) injury is a rare but severe complication of thyroid surgery. A surgeon’s experience and knowledge about anatomy is the keystone of preventing RLN injury [1]. Identifying the RLN visually during surgery can lower the incidence of RLN injury and the permanent injury rate to 0.9% [2–4].

Thyroid surgery induces adjacent tissue scarring and distortion of the anatomy. The distorted anatomy increases the incidence of the RLN injury during subsequent thyroid surgery, even if done by a high-volume surgeon (2% to 12%) [5–7].

Intraoperative nerve monitoring (IONM) is a technique that can help identify or map the course of the RLN nerve and recognize its function [1]. One study [8] reported that in the general population, using IONM provided no significant benefit for preventing permanent RLN injury. However, in some high-risk patients, such as in a reoperation or surgery for malignancy, using IONM might be beneficial, although this is still controversial [1, 9–11].

In previous studies [9–11], patients with a reoperation were included, and surgeons counted the total number of nerves during the exploration in the second operation. However, none of these studies definitely pointed out the actual RLN injury rate during the re-exploration. These studies may in their reoperation denominator an RLN that had never been explored before, which would underestimate the actual injury rate during every re-exploration. Also, none of these studies clearly compared the RLN injury rate during re-exploration between groups in which IONM was and was not used. In addition, data on the incidence of RLN injuries in re-explorations done by a single high-volume surgeon before and after the introduction of IONM are lacking.

In this study, we focus on the effect of IONM in the re-exploration of the RLN and the experience of a single high-volume surgeon.

Methods

Definitions

‘Reoperation’ means a second or subsequent surgery to correct the same condition or complications of a previous surgery. ‘Re-exploration’ means a second or subsequent nerve exploration for a reoperation.

Material

An IONM device was introduced to our institute in March 2006. All patients given thyroid reoperations with IONM were included in this study. All the medical records were reviewed retrospectively for gender, age, diagnosis, indications of the first and second surgeries, operative methods of the first and second surgeries, time interval between the two operations, and the signs and symptoms of RLN paresis before and after reoperations. A preoperative laryngoscopy exam was routinely done by an otolaryngologist for every patient given a reoperation.

All reoperations were done by a single high-volume surgeon who had already done more than 4000 thyroid surgeries before we conducted this study. All patients given a reoperation between 2001 and 2006 by the same surgeon without IONM were also reviewed.

Exclusion criteria included unknown methods for the previous surgery and patients who had not had an RLN re-exploration during the reoperation. Patients who had vocal cord palsy from a previous operation, detected in a laryngoscope examination, were also excluded.

Between 2001 and 2010, 210 patients were reoperated on by the surgeon. Only 56 patients given an RLN re-exploration using IONM and 15 patients without IONM were included in this study.

IONM

The IONM device was set up in a standard manner with two electrodes on the endotracheal tube (NIM EMG Endotracheal Tube; Medtronic, Minneapolis, MN, USA). A four-step procedure was used to ensure that the device was functioning adequately (Table 1) [12, 13].

Table 1 Four-step intraoperative neuromonitoring (IONM) procedure during thyroid surgery

All patients received transverse cervical incision during reoperation. Lateral approach or medial approach is determined by the result of preoperative sonography performed by the surgeon. We routinely checked the vagus nerve, mapped the course of the RLN, and always did IONM stimulation tests before we ligated or transected any tissue during surgery. The stimulation current is set at 1 mA. However, 2 mA is used sometimes to map the route of RLN while not seen very clearly. The normal range of amplitude is 700 to 1500 μv.

Statistical method

We used Fisher’s exact test for comparing the two-tailed P value between the two groups. Significance was set at P <0.05.

Results

Fifty-six patients with a definite re-exploration of the RLN were analyzed. IONM was used to re-explore 70 RLNs. The indications and operative methods of the first and second operations are shown in Table 2.

Table 2 Re-exploration of the recurrent laryngeal nerve (RLN) with and without intraoperative neuromonitoring (IONM)

Of the 70 RLNs re-explored with IONM, three were transected (Table 3): two nerves were intentionally transected and one (1.43%) was accidentally injured. In comparison, of the 15 RLNs re-explored without IONM, three (20%, P = 0.0164) nerves were accidentally injured (Tables 2 and 4). The overall injury rate of RLNs during reoperation for this surgeon was 1.90% (4 RLNs in 210 cases).

Table 3 Information on permanent injury of the recurrent laryngeal nerve (RLN) using intraoperative neuromonitoring (IONM) ( n = 3; 5.36% per patient; 4.29% per nerve and 1.43% of incidental injury, P = 0.0164)
Table 4 Information on permanent injury of the recurrent laryngeal nerve (RLN) without intraoperative neuromonitoring (IONM) ( n = 3, 20% per patient and nerve)

We routinely checked the vagus nerve, mapped the course of the RLN, and always did IONM stimulation tests before we ligated or transected any tissue during surgery. However, the IONM device malfunctioned once during an operation, which resulted in an accidental RLN transection. We were unaware of the RLN injury until we used the IONM to check the vagus nerve and received no response. In contrast, the contralateral vagus nerve responded well to stimulation. We identified the transected end of the RLN using a nerve stimulator and did a re-anastomosis with the ansa-cervicalis nerve.

In another patient, one RLN was transected because of a tumor encasement. The short residual nerve end beneath Berry’s ligament was confirmed using a nerve stimulation test and was re-anastomosed with the ansa-cervicalis nerve. The patient had an acceptable voice during follow-up despite laryngoscope-revealed vocal cord palsy.

In a third patient, the RLN was intentionally transected because of recurrent thyroid cancer with tracheal involvement. We initially tried to do a segmental tracheal resection and primary anastomosis to preserve her larynx. However, the RLN lost function during surgery even though the nerve appeared to be visually intact. The injury might have been caused by excessive traction during dissection, but we could not tell whether it was a temporary or permanent injury. For safety, we changed the operation to a total laryngectomy and sacrificed the RLN as well.

Discussion

In reoperative thyroid surgery, there is higher rate of complications because adjacent tissue is scarred and anatomy is distorted. The scarring makes dissection more difficult, and bleeding further compromises the operation field. Distorted anatomy increases the injury rate of important adjacent structures [1]. Even for an experienced surgeon, reoperative surgery is stressful.

RLN injury in reoperative surgery occurs primarily because of difficulty in identifying the RLN. Identifying the RLN lowers the incidence of permanent RLN injury from 5.2% to 1.2% [4]. In addition, the RLN injury rate in reoperative surgery is around 1% to 12% in different studies [5–7]. IONM helps to identify and map the route of the RLN during reoperation, but whether it prevents RLN injury is still controversial [1, 9–11]. One study [10] showed that the rate of RLN injury was higher in a group of reoperative thyroidectomies done without IONM (19%) than in a group done with IONM (7.8%). Another study [1] reported 1.9% and 1.7% permanent RLN injury rates for reoperative thyroidectomies with and without IONM, respectively [1]. However, no study presents the actual RLN injury rate during reoperative thyroid surgery with RLN re-exploration, nor does any show that IONM is beneficial for re-exploring the RLN.

Another study [14] includes 89 and 157 patients with and without IONM usage, respectively. And the incidence of transient nerve palsy rate is higher in the group with IONM usage (6.2 vs. 2.5%), but showed no statistically significance. The incidence of permanent injury rate is 0% in the group with IONM usage, and 0.6% in those without. In this study, <10% of the patient was diagnosed as thyroid malignancy and the operation is performed by eight surgeons with different experience. Besides, the patient group is selected according to the surgeon’s preference, which would cause huge selection bias. In our study, patients with IONM usage showed significant benefit in preventing RLN injury and the reason may be that we included more thyroid malignancy in our patient group. In our study, 42.9% of patients with IONM and 80% of patients without IONM are diagnosed with thyroid malignancy and this may increase the difficulty of reoperation due to tumor invasion. All the operations in our study are performed by a single high-volume surgeon and all patients received IONM usage after introduction of IONM, which would significantly reduce bias from patient selection and surgeon’s technique.

In our study, both groups of patients had a definite history of exploration and re-exploration of RLNs. The actual injury rate of each re-exploration was compared. For the group without IONM, the accidental injury rate was 20% (3 of 15 nerves), and for the group with IONM, it was a significantly (P <0.05) different 1.43% (1 of 70 nerves). The overall reoperation injury rate in our institute is 1.90% (4 of 210 cases). Although the benefit of using IONM for reoperative thyroid surgery may not be indisputable, we hypothesize that if re-exploration near RLNs is planned and done using IONM, significantly fewer accidental RLN injuries will occur. IONM will also give surgeons more confidence when re-exploring RLNs, as shown with the patients in this study.

Several potential pitfalls of IONM have been reported [15, 16], such as device malfunction, improper device setup, misuse of muscle relaxant, anatomic variation of the RLN, and shunt stimulus, which can cause misleading information. Device set-up problems are the most common cause of false IONM results. In the one case of accidental injury in this study, an RLN was transected because the monitor temporarily stopped functioning during surgery. To prevent this from occurring, a standard manual for device set-up and proper endotracheal tube insertion is necessary [13]. However, the possibility that the IONM device will malfunction or require readjustment during the operation still exists. Intermittently stimulating the vagus nerve using a recently designed anchor electrode [17, 18] in addition to a conventional handheld bipolar stimulation electrode during surgery may alert the surgeon to device malfunction. In our first case of RLN injury, device malfunction was identified as the main reason of injury. Continuous IONM would gain a lot of benefit in this situation and prevent nerve injury [19].

There are other advantages of using IONM during thyroid surgery. For accidental transection or deliberate sacrifice of the RLN, IONM helps when searching for the transected end. Once the transected end is identified, better re-anastomosis of the RLN can be achieved. In addition, IONM helps identify the transected end of RLNs with very short ends and RLNs that are buried in the Berry’s ligament. This should allow a more precise nerve repair or even re-anastomosis with the ansa cervicalis. However, the actual benefits require more investigation [20]. IONM revealed advantage in endoscopic surgery in animal study, but the benefit in human need further investigation [21].

For patients with bilateral disease or preoperative unilateral RLN paresis, IONM may guide the strategy of the thyroid operation. IONM can be used to monitor the RLN function because that function may still be impaired even though the RLN visually appears to be intact. In one study [22], 33 patients developed unilateral cord paralysis, but only five nerve injuries were recognized during surgery. In another [3], 40 patients developed cord paralysis with only three recognizable injuries. The possible mechanisms include forceps clamping, nerve stretching, electrothermal injury, ligature entrapment, and ischemia. A stretching injury may be the main reason of temporary paresis [12]. If a temporary or permanent nerve dysfunction occurs during the operation, the surgical strategy may be changed for safety [23]. For patients who already have unilateral vocal cord paresis, an operation on the contralateral side using IONM will help, not to prevent an RLN injury, but to detect RLN paresis, to guide postoperative care, or even to prompt the surgeon to do a protective tracheostomy after the thyroid surgery. One of our patients developed RLN paresis during the operation and we changed our strategy to a total laryngectomy with a tracheostomy.

This study has some limitations. The first is the lack of randomized control groups because we have used an IONM device for every patient given a reoperation since it was introduced. Therefore, we used as our control group patients operated on before the introduction of the IONM. It is almost impossible to do a prospective randomized control study for this topic for three reasons. First, there are not many such cases every year. Second, if we select cases from several surgeons, differences in technique will significantly bias the results. Third, it is unethical to put high-risk patients into a control group that does not allow the surgeon to use currently available technology to help identify the route of the nerve so that it will not be accidently damaged. Therefore, this study presents the usefulness of IONM during thyroid surgery done by a single high-volume surgeon. The second is that we did more reoperation after introduce of IONM since 2006 (15 cases in 5 years vs. 56 cases in 5 years). Before introduction of the device, we reoperated in a limited indication. And the percentage of thyroid cancer is higher in the group without IONM usage. After IONM usage, we must say that we gain more assurance and widen the indication of reoperation to recurrent benign goiter or Graves’ disease. In our experience, avoiding direct dissection over the RLNs is the best way to prevent injury. The IONM helps to identify or map the route of RLNs, but it does not always prevent the surgeon from transecting the nerve.

Conclusion

This is the first paper that clearly states the influence of IONM on a single high-volume surgeon during re-exploration of the RLN. Using IONM for re-exploring RLNs lowers the nerve-injury rate. In addition, it may help guide surgical strategy and even help with nerve re-anastomosis. Therefore, we recommend using IONM for this high-risk group of patients.

Abbreviations

IONM:

Intra-operative nerve monitoring

RLN:

Recurrent laryngeal nerve.

References

  1. Yarbrough DE, Thompson GB, Kasperbauer JL, Kasperbauer JL, Harper CM, Grant CS: Intraoperative electromyographic monitoring of the recurrent laryngeal nerve in reoperative thyroid and parathyroid surgery. Surgery. 2004, 136: 1107-1115. 10.1016/j.surg.2004.06.040.

    Article  PubMed  Google Scholar 

  2. Lahey FH: Routine dissection and demonstration of recurrent laryngeal nerves in subtotal thyroidectomy. Surg Gynecol Obstet. 1938, 66: 775-777.

    Google Scholar 

  3. Chiang FY, Wang LF, Huang YF, Lee KW, Kuo WR: Recurrent laryngeal nerve palsy after thyroidectomy with routine identification of the recurrent laryngeal nerve. Surgery. 2005, 137: 342-347. 10.1016/j.surg.2004.09.008.

    Article  PubMed  Google Scholar 

  4. Jatzko GR, Lisborg PH, Muller MG, Wette VM: Recurrent nerve palsy after thyroid operations–principal nerve identification and a literature review. Surgery. 1994, 115: 139-144.

    CAS  PubMed  Google Scholar 

  5. Tissel L, Hansson G, Jansson S, Salander H: Reoperation in the treatment of asymptomatic metastasizing medullary thyroid carcinoma. Surgery. 1986, 99: 60-66.

    Google Scholar 

  6. Goretzki P, Simon D, Frilling A, Witte J, Reiners C, Grussendorf M, Horster FA, Roher HD: Surgical reintervention for differentiated thyroid cancer. Br J Surg. 1993, 80: 1009-1012. 10.1002/bjs.1800800826.

    Article  CAS  PubMed  Google Scholar 

  7. Patou CA, Norton JA, Brennan MF: Hypocalcemia following thyroid surgery: incidence and prediction of outcome. World J Surg. 1998, 22: 718-724. 10.1007/s002689900459.

    Article  Google Scholar 

  8. Barczynski M, Konturek A, Cichon S: Randomized clinical trial of visualization versus neuromonitoring of recurrent laryngeal nerves during thyroidectomy. Br J Surg. 2009, 96: 240-246. 10.1002/bjs.6417.

    Article  CAS  PubMed  Google Scholar 

  9. Atallah I, Dupret A, Carpentier AS, Weingertner AS, Volkmar PP, Roder JF: Role of intraoperative neuromonitoring of the recurrent laryngeal nerve in high-risk thyroid surgery. J Otolaryngol Head Neck Surg. 2009, 38: 613-618.

    PubMed  Google Scholar 

  10. Chan WF, Lang BH, Lo CY: The role of intraoperative neuromonitoring of recurrent laryngeal nerve during thyroidectomy: a comparative study on 1000 nerves at risk. Surgery. 2006, 140: 866-872. 10.1016/j.surg.2006.07.017.

    Article  PubMed  Google Scholar 

  11. Thomusch O, Sekulla C, Timmermann W, Neumann HJ, Kruse E, Muhlig HP, Hamelmann WH, Richter CH, Voss J, Dralle H: Intraoperative neuromonitoring in thyroid surgery. Results of the German Prospective Multicentre Study. Eur Surg. 2003, 35: 240-245. 10.1007/s10353-003-0019-6.

    Article  Google Scholar 

  12. Chiang FY, Lu IC, Kuo WR, Lee KW, Chang NC, Wu CW: The mechanism of recurrent laryngeal nerve injury during thyroid surgery–the application of intraoperative neuromonitoring. Surgery. 2008, 143: 743-749. 10.1016/j.surg.2008.02.006.

    Article  PubMed  Google Scholar 

  13. Chiang FY, Lee KW, Chen HC, Chen HY, Lu IC, Kuo WR, Hsieh MC, Wu CW: Standardization of intraoperative neuromonitoring of recurrent laryngeal nerve in thyroid operation. World J Surg. 2010, 34: 223-229. 10.1007/s00268-009-0316-8.

    Article  PubMed  Google Scholar 

  14. Alesina P, Rolfs T, Hommeltenberg S, Hinrichs J, Meier B, Mohmand W, Hofmeister S, Walz MK: Intraoperative neuromonitoring does not reduce the incidence of recurrent laryngeal nerve palsy in thyroid reoperations: results of a retrospective comparative analysis. World J Surg. 2012, 36: 1348-1353. 10.1007/s00268-012-1548-6.

    Article  PubMed  Google Scholar 

  15. Beneragama T, Serpell JW: Extralaryngeal bifurcation of the recurrent laryngeal nerve: a common variation. ANZ J Surg. 2006, 76: 928-931. 10.1111/j.1445-2197.2006.03899.x.

    Article  PubMed  Google Scholar 

  16. Katz AD: Extralaryngeal division of the recurrent laryngeal nerve. Report on 400 patients and the 721 nerves measured. Am J Surg. 1986, 152: 407-410. 10.1016/0002-9610(86)90313-2.

    Article  CAS  PubMed  Google Scholar 

  17. Schneider R, Przybyl J, Hermann M, Hauss J, Jonas S, Leinung S: A new anchor electrode design for continuous neuromonitoring of the recurrent laryngeal nerve by vagal nerve stimulations. Langenbecks Arch Surg. 2009, 394: 903-910. 10.1007/s00423-009-0503-y.

    Article  PubMed  Google Scholar 

  18. Ulmer C, Koch KP, Seimer A, Molnar V, Meyding-Lamade U, Thon KP, Lamade W: Real-time monitoring of the recurrent laryngeal nerve: an observational clinical trial. Surgery. 2008, 143: 359-365. 10.1016/j.surg.2007.10.007.

    Article  PubMed  Google Scholar 

  19. Friedrich C, Ulmer C, Rieber F, Kern E, Kohler A, Schymik K, Thon KP, Lamade W: Safety analysis of vagal nerve stimulation for continuous nerve monitoring during thyroid surgery. Laryngoscope. 2012, 122: 1979-1987. 10.1002/lary.23411.

    Article  PubMed  Google Scholar 

  20. Miyauchi A, Inoue H, Tomoda C, Fukushima M, Kihara M, Higashiyama T, Takamura Y, Ito Y, Kobayashi K, Miya A: Improvement in phonation after reconstruction of the recurrent laryngeal nerve in patients with thyroid cancer invading the nerve. Surgery. 2009, 146: 1056-1062. 10.1016/j.surg.2009.09.018.

    Article  PubMed  Google Scholar 

  21. Witzel K, Benhidjeb T: Monitoring of the recurrent laryngeal nerve in totally endoscopic thyroid surgery. Eur Surg Res. 2009, 43: 72-76. 10.1159/000220596.

    Article  CAS  PubMed  Google Scholar 

  22. Lo CY, Kwok KF, Yuen PW: A prospective evaluation of recurrent laryngeal nerve paralysis during thyroidectomy. Arch Surg. 2000, 135: 204-207. 10.1001/archsurg.135.2.204.

    Article  CAS  PubMed  Google Scholar 

  23. Goretzki PE, Schwarz K, Brinkmann J, Wirowski D, Lammers BJ: The impact of intraoperative neuromonitoring (IONM) on surgical strategy in bilateral thyroid diseases: is it worth the effort?. World J Surg. 2010, 34: 1274-1284. 10.1007/s00268-009-0353-3.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Shih-Ming Huang.

Additional information

Competing interests

The authors declare they have no competing interests.

Authors’ contributions

YC participated in the design of the study and performed the statistical analysis, participated in the sequence alignment, and drafted the manuscript. SH conceived of the study, and participated in its design and coordination and helped to draft the manuscript. Both authors read and approved the final manuscript.

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Chuang, YC., Huang, SM. Protective effect of intraoperative nerve monitoring against recurrent laryngeal nerve injury during re-exploration of the thyroid. World J Surg Onc 11, 94 (2013). https://doi.org/10.1186/1477-7819-11-94

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1477-7819-11-94

Keywords